Investigating Drip Infusion Speed Control: A Study of Weight and Pinch Methods Concerning Subject Variations

Lanny Agustine (1) , Made Indra Ayu Astarini (2) , Maria Manungkalit (3) , Hartono Pranjoto (4)
(1) Department of Electrical Engineering, Widya Mandala Surabaya Catholic University Kalijudan 37 Street, Surabaya, East Java, Indonesia , Brazil
(2) Nursing Study Program, Faculty of Nursing, Widya Mandala Surabaya Catholic University Kalisari Selatan 1 Street, Pakuwon City, Surabaya, East Java, Indonesia , Indonesia
(3) List all distinct addresses in the same way , Indonesia
(4) Department of Electrical Engineering, Widya Mandala Surabaya Catholic University , Indonesia

Abstract

An infusion drip system that utilizes a clamp method to control the drip rate has not yet achieved a consistent drip rate across different subjects. The condition was that each movement of the pinch valve did not consistently stimulate the release of one drop of liquid. In this study, tests were conducted with variations in the pinch-release period (3 seconds, 2.5 seconds, 2 seconds, and 1.5 seconds) and variations in the release pinch valve duration (55 milliseconds, 60 milliseconds, 65 milliseconds, 70 milliseconds, and 75 milliseconds). The testing involved 15 participants, including 100 cc of NaCl infusion fluid for 1-2 hours. The results indicated two variations in drip speed among all test subjects: it required either two ticks to produce one drop of infusion (the dominant outcome) or just one tick. Based on tests with healthy participants and their responses to questionnaires, it was concluded that the device was quite comfortable to use with the release pinch valve duration of 60-70 milliseconds and a pinch-release period of 2 seconds, 2.5 seconds, or 3 seconds.

References

[1] J. Silva, L. Gonçalves, and P. P. Sousa, Fluid therapy and shock: an integrative literature review, Br. J. Nurs. 27, 8, 449–454 (2018)
[2] R. C. L. da Silva, T. QuinellatoLouro, A. A. de F. Peregrino, C. R. L. da Silva, C. B. Marta, and A. Itria, Cost-effectiveness of infusion pumps to reduce errors in a Pediatric ICU, Rev. Bras. Enferm. 72, 3, 617–623 (2019)
[3] S. Padhi, I. Bullock, L. Li, and M. Stroud, Intravenous fluid therapy for adults in hospital: summary of NICE guidance, Br. Med. J. 347, dec10 1, f7073–f7073 (2013)
[4] A. P. A. Moreira, C. L. Escudeiro, B. P. Christovam, Z. R. Silvino, M. F. de Carvalho, and R. C. L. da Silva, Use of technologies in intravenous therapy: contributions to a safer practice, Rev. Bras. Enferm. 70, 3, 595–601 (2017)
[5] D. Oros, M. Penˇci´c, M., J. Šulc, M. Cavi´c, S. Stankovski, G. Ostoji´c, O. Ivanov, Smart intravenous infusion dosing system, Appl. Sci., 11, 2, 513 (2021)
[6] G. M. Eastwood et al., Intravenous fluid administration and monitoring for adult ward patients in a teaching hospital, Nurs. Heal. Sci. 14, 2, 265–271 (2012)
[7] R. F. R. McCrory, G. J. Gormley, A. P. Maxwell, and T. Dornan, Learning to prescribe intravenous fluids: A scoping review, Perspect. Med. Educ. 6, 6, 369–379 (2017)
[8] Jaganathan, A. Kumaran, J. Sujitha, Janani, Mathumitha, Smart Intravenous Drip Infusion Monitoring and Health Information System, Int. J. of Adv. in Eng. and Manage. (IJAEM), 5, 4, 493-500 (2023)
[9] O. Ajibola, O. Sunday, D. Eyehorua, Development of Automated Intravenous Blood Infusion Monitoring System using Load Cell Sensor, J. Appl. Sci. Environ. Manage. 22, 10, 1557–1561 (2018)
[10] M. Jindal, N. Gajjar, and N. Patel, IoT-Based Smart Intravenous Drip Monitoring System, Adv. in Intel. Sys. and Comp. (AISC), 1299, 463–468 (2021)
[11] D. Lestariningsih, H. Pranjoto, L. Agustine, Y. D. W. Werdani, and B. Teja, Load Cell Application for Infusion Fluid Volume Monitoring System, Saintek J. Res. 26, 2, 165–177 (2021) (in Bahasa).
[12] K. N. T. Yayer, W. A. Weliamto, R. Sitepu, and H. Pranjoto, Monitoring and Termination of Infusion Fluids Using Digital Infusion Scales By Utilizing Web Server, Simetris J. Mech. Electr. Comput. Sci. Eng. 11, 1, 55–64 (2020) (in Bahasa).
[13] L. Agustine, M. I. A. Astarini, M. Manungkalit, J. Amadeus, and H. Pranjoto, The impact assessment of automated drip infusion control using weighing scale and pinch method on subjects, InCASST 2023, E3S Web of Conf. 475, 02002 (2024)
[14] M. R. S. Zarandi and H. Khoshab, A Comprehensive Formula for Calculating the Infusion Rate of Medications Diluted with Infusion Solution, Open Nurs. J. 13, 1, 136–138 (2019)
[15] F. H. Zhou, C. Liu, Z. Mao, and P. L. Ma, Normal saline for intravenous fluid therapy in critically ill patients, Chinese J. Traumatol. 21, 1, 11–15(2018)

Authors

Lanny Agustine
Made Indra Ayu Astarini
Maria Manungkalit
Hartono Pranjoto
[1]
“Investigating Drip Infusion Speed Control: A Study of Weight and Pinch Methods Concerning Subject Variations”, Soc. sci. humanities j., vol. 9, no. 01, pp. 6394–6400, Jan. 2025, doi: 10.18535/sshj.v9i01.1449.